Hydroxide Ion-Selective Polymeric Membrane-Coated Wire Electrode Based on Oxomolybdenum(V) Tetraphenylporphyrin Complex

  • Hideo Abe
    Insitute of Applied Biochemistry, University of Tsukuba
  • Etsuo Kokufuta
    Insitute of Applied Biochemistry, University of Tsukuba

抄録

<jats:title>Abstract</jats:title> <jats:p>A coated wire electrode (CWE) was prepared by use of a poly(vinyl chloride)-supported liquid membrane containing oxoethoxo(5,10,15,20-tetraphenylporphyrinato)molybdenum(V) as a carrier. CWE based on such a metallo-porphyrin complex exhibits near-Nernstian responses to varying concentrations of NaOH, NaF, and C2H5OH, especially to NaOH over a wide concentration range. The pOH response curve was influenced by certain anions other than OH− ion, but not by cations such as Li+, Na+, K+, and Cs+. The mixed solution method gave a selectivity sequence OH−&gt;&gt;ClO4−&gt;NO2−&gt;F−∼SCN−&gt;I−&gt;NO3−&gt;SO3−&gt;Br−&gt;SO42−, with a preference for OH− over ClO4− by 104.6 and over F− (or SCN−) by 106.2. In addition, Cl− ion was found to have little interfering effect under conditions used in the measurements of selectivity coefficients. Both response-time profile and pOH-response curve scarcely varied even when CWE was stored in distilled water for 167 days.</jats:p>