インピーダンス・オージオメトリ

  • 神崎 仁
    国際医療福祉大学熱海病院耳鼻咽喉科

書誌事項

タイトル別名
  • Impedance Audiometry

抄録

Herein, the author briefly describes impedance audiometry, which includes tympanometry and evaluation of the acoustic reflex. Tympanometry is useful for the evaluation of pathological conditions of the tympanic membrane and middle ear. Tympanometry is applied routinely for the screening of hearing in neonates, along with evaluation of the auditory brainstem response (ABR). It has, however, recently been reported that tympanometry using 1000Hz as the probe tone in neonates younger than 3 months of age is easier to interpret and more reliable than 226-Hz tympanometry. Two methods to test the acoustic reflex (ipsi- and contralateral stimulation) can be used, and the test is still useful for evaluating the acoustic reflex pathway including the VIIth and VIIIth nerves in the brainstem in situations where the ABR and/or otoacoustic emission tests are not available.