肺定位放射線照射における計算精度―計算アルゴリズムの比較―

書誌事項

タイトル別名
  • ACCURACY OF CALCULATIONS FOR STEREOTACTIC RADIOTHERAPY OF THE LUNG-COMPARISON OF DOSE CALCULATION ALGORITHMS-
  • COMPARISON OF DOSE CALCULATION ALGORITHMS
  • 計算アルゴリズムの比較

この論文をさがす

抄録

The purpose of this study is to evaluate the accuracy of dose calculations by three algorithms.<BR>Methods and Materials: Depth dose, OPF (Output Factor) and dose profiles were measured in a heterogeneous phantom. These values were also calculated by three algorithms of the Batho power law (BPL), Equivalent-TAR (ETAR) and Convolution superposition (CS). The data were obtained for 4, 6 and 10 MV photon beams with a linear accelerator (Varian 21EX). Field size ranged from 3-3cm2 to 10-10cm2. Dose profiles of beam penumbra were also measured by a 0.125ml ionization chamber at the point of 8, 13 and 18cm from the surface of the phantom at intervals of 1mm.<BR>Result: Differences between measured and calculated depth doses were within 2% in BPL and CS, but depth doses were overestimated in ETAR. OPFs were also overestimated with the error of more than 4% in ETAR. Absorbed dose calculated by CS were in agreement with the values measured by the ionization chamber. The absorbed dose by BPL and ETAR, however, were overestimated in particular in absorbed dose of 85-95%. Conclusion: Although isodose lines by CS are precisely correct, the measured absorbed dose may be smaller than the prescribed doses calculated by BPL and ETAR. It should be noticed that calculated isodose lines may be significantly different from measured values in the treatment of lung tumors.

収録刊行物