BNP測定試薬の基礎的検討およびBNP値と心エコー検査所見との関連について

書誌事項

タイトル別名
  • Fundamental assessment of chemiluminescent enzyme immunoassay measurement reagent “BNP-JP: Architect”and the relationship between plasma concentrations of brain natriuretic peptide (BNP) and parameters of echocardiography

この論文をさがす

抄録

We examined the fundamental performance of a reagent “BNP-JP Architect” and the relationship between the concentrations of brain natriuretic peptide (BNP) and the parameters of echocardiography. BNP-JP Architect had satisfactory performance, except for the effect of hemolysis. The relationship between plasma BNP level and the parameters of echocardiography was evaluated in 120 patients with sinus rhythm (mean age 69 ± 13 years old, 70 men, 50 women). Plasma BNP level correlated with left atrial dimension (LAD) (r = 0.39, p < 0.01), left ventricular end-diastolic dimension (r = 0.22, p < 0.05), ejection fraction (EF) (r = –0.40, p < 0.01), and ratio of the peak velocity of transmitral flow during early diastole to the peak myocardial velocity during early diastole at septal annulus (E/e') (r = 0.53, p < 0.01). In multivariate linear regression analysis, EF (p = 0.04) and E/e' (p < 0.01) were independent determinants of plasma BNP level. Patients with decreased renal function (estimated glomerular filtration ratio (eGFR) <60 ml/min/1.73 m2) were older (p < 0.01) and had higher LAD (p < 0.05), E/e' (p = 0.01), and plasma BNP level (p < 0.01) than patients with preserved renal function (eGFR≧60 ml/min/1.73 m2). Although renal function should be considered when interpreting plasma BNP level, the plasma BNP level measured by BNP-JP Architect can be used as a surrogate marker of left ventricular (LV) systolic function and LV filling pressure and provide useful information for the diagnosis of heart failure.

収録刊行物

  • 医学検査

    医学検査 63 (1), 29-35, 2014

    一般社団法人 日本臨床衛生検査技師会

参考文献 (6)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘

ページトップへ