体内ロボット用非接触給電における周波数と給電領域の検討 Examination of Feeding Range and Frequency for Wireless Power Transfer for Robot inside the Body

この論文にアクセスする

この論文をさがす

抄録

Research on the robot inside the body equipped with various features, such as therapeutic function and self-propelled capability within the capsule endoscope is performed. Research on Wireless Power Transfer for robot inside the body to act sufficient in the body is performed, because internal battery has a limit of the capacity. It is necessary to reduce an electromagnetic field because electromagnetic field is generated from a Wireless Power Transfer. Power transmission in the Wireless Power Transfer depends largely on the quality factor of the coil and the transmission frequency. In this paper, it is shown that it is possible transmission power of 30 mW in the range of 300 mm × 300 mm × 180 mm, while reducing an electromagnetic field in the frequency 700 kHz.

収録刊行物

  • 日本AEM学会誌

    日本AEM学会誌 23(1), 144-150, 2015

    The Japan Society of Applied Electromagnetics and Mechanics

各種コード

  • NII論文ID(NAID)
    130005073647
  • NII書誌ID(NCID)
    AN10457520
  • 本文言語コード
    EN
  • ISSN
    0919-4452
  • NDL 記事登録ID
    026346812
  • NDL 請求記号
    Z15-B3
  • データ提供元
    NDL  J-STAGE 