Radial Profile Measurements of Electron Temperature and Density Using the Thomson Scattering System in GAMMA 10/PDX

この論文をさがす

抄録

<p>We developed the YAG Thomson scattering (TS) system for electron temperature and density radial profiles in a single laser shot in the tandem mirror GAMMA 10/PDX. The optical collection system for the TS light is constructed using spherical mirrors and nine bundled optical fibers. The measurable radial positions are 5-cm intervals in a region of ±20 cm due to the fixed position of the bundled optical fibers, which are set on a fixed-height optical platform. We replaced this with a lab jack to make the bundled optical fiber positions movable. Consequently, we can measure the radial profiles of the electron temperature in more detail. Moreover, we added a third spherical light collection mirror at the bottom of the main spherical collection mirror to increase the TS light intensity in the edge plasma region. In this way, we can obtain a slightly larger TS signal intensity than in the case without a third collection mirror. The radial electron temperatures and densities at 15 radial positions in GAMMA 10/PDX are successfully obtained. For the in-situ calibration of the electron density measurements of the TS system, we compare the TS system measurements to the electron line density measured using a microwave interferometer system.</p>

収録刊行物

  • Plasma and Fusion Research

    Plasma and Fusion Research 13 (0), 3402051-3402051, 2018-05-10

    一般社団法人 プラズマ・核融合学会

参考文献 (17)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ