Time-of-Flight Mass Spectrometer with Improved Resolution

  • W. C. Wiley
    Bendix Aviation Corporation Research Laboratories, Detroit, Michigan
  • I. H. McLaren
    Bendix Aviation Corporation Research Laboratories, Detroit, Michigan

Abstract

<jats:p>A new type of ion gun is described which greatly improves the resolution of a nonmagnetic time-of-flight mass spectrometer. The focusing action of this gun is discussed and analyzed mathematically. The validity of the analysis and the practicability of the gun are demonstrated by the spectra obtained. The spectrometer is capable of measuring the relative abundance of adjacent masses well beyond 100 amu.</jats:p>

Journal

Citations (119)*help

See more

Keywords

Details 詳細情報について

Report a problem