Pressure Measurement Made by the Utilization of Ruby Sharp-Line Luminescence

この論文をさがす

抄録

<jats:p> A rapid, convenient technique for precision pressure measurement in the diamond-anvil high-pressure cell, which makes use of the sharp-line (R-line) luminescence of ruby, has been developed. The observed shift is -0.77 ± 0.03 reciprocal centimeters per kilobar for R <jats:sub>1</jats:sub> and -0.84 ± 0.03 reciprocal centimeters per kilobar for R <jats:sub>2</jats:sub> to lower energy and is approximately linear in the range studied (to 22 kilobars). Line-broadening has been observed in some instances and has been tentatively identified with nonhydrostatic conditions surrounding the ruby sample. </jats:p>

収録刊行物

  • Science

    Science 176 (4032), 284-285, 1972-04-21

    American Association for the Advancement of Science (AAAS)

被引用文献 (17)*注記