Retrieval techniques for managing flexible intracoronary stent misplacement
抄録
<jats:title>Abstract</jats:title><jats:p>With the increasing use of flexible intracoronary stents, the likelihood of complications, including stent misplacement, will tend to rise. We describe the successful use of three commercially available retrieval devices: the nitinol gooseneck snare, the biliary forceps, and the multipurpose basket. We recommend the availability of these devices to operators involved in intracoronary stent placement. © 1993 Wiiey‐Liss, Inc.</jats:p>