抄録

Elution tests with pure water (PW), 1N nitric acid (NA), 544ppm sodium hypochlorite and bovine plasma (BP) were performed to elucidate the elution of membrane constituents for ceramic membranes based on the Standards for Approval of Hemodialyzers, the Ministry of Health and Welfare, Japan, Sintered alumina ceramic membranes elute a fairly small amounts of membrane constituents for any elution tests. Porous glass membranes are detected no elution of aluminum, and a large amounts of boron and silicon with any solvent. After rinsing with NA, B and Si are eluted in small amounts with PW and in large amounts with BP. Surface modification is required to prevent the elution of boron and silicon for porous glass membranes.