Specific volume measurements of Cu, Mo, Pd, and Ag and calibration of the ruby <i>R</i>1 fluorescence pressure gauge from 0.06 to 1 Mbar
-
- H. K. Mao
- Geophysical Laboratory, Carnegie Institution of Washington, Washington, D.C. 20008
-
- P. M. Bell
- Geophysical Laboratory, Carnegie Institution of Washington, Washington, D.C. 20008
-
- J. W. Shaner
- Lawrence Livermore Laboratory, University of California, Livermore, California 94550
-
- D. J. Steinberg
- Lawrence Livermore Laboratory, University of California, Livermore, California 94550
抄録
<jats:p>The wavelength shift with pressure of the ruby R1 fluorescence line (Δλ) has been calibrated in the diamond-window pressure cell from 0.06 to 1 Mbar. This was done by simultaneously making specific volume measurements of four metals (Cu, Mo, Ag, and Pd) and referring these results to isothermal equations of state derived from shock-wave experiments. The result is P (Mbar) = (19.04/5) {[(λ0+Δλ)/λ0]5−1}, where λ0 is the wavelength measured at 1 bar.</jats:p>