Fe<SUP>2+</SUP>/Fe<SUP>3+</SUP> Mapping with an Electron Probe Microanalyzer (EPMA)

  • KATO Takenori
    Department of Earth and Planetary Sciences, Graduate School of Science, Nagoya University

Bibliographic Information

Other Title
  • 電子線プローブマイクロアナライザー (EPMA) によるFe<SUP>2+</SUP>/Fe<SUP>3+</SUP>値マッピング
  • デンシセン プローブマイクロアナライザー EPMA ニ ヨル Fe2 Fe3 チ マッピング

Search this article

Abstract

The relative intensities of Fe Lα1, 2 and Lβ1 X-ray emission peaks differ significantly with valance state and bond association. These X-ray emission peaks cannot be distinguished from each other using an energy-dispersive spectrometer (EDS), because the difference in energy between these peaks is smaller than the spectral resolution of EDS (approximately 100-150 eV). However, these emission peaks can be distinguished from each other using a wavelength-dispersive spectrometer (WDS). A crystal whose 2d value is about 60 A is considered to be suitable for analyzing the first order of Fe Lα1, 2 and L β1 X-ray emission peaks. The second order of these peaks can also be analyzed using lead stearate (STE : 2d=100.4 Å) crystal. Although it is difficult to use the ' peak seek' routine for a mapping analysis, the effects of chemical shifts can be suppressed by configuring the spectrometer. The interferences of high orders of Si and Na emission peaks can be suppressed by a pulse height analysis. Applying a mapping technique to measuring Fe Lα 1, 2 and Lβ1 emission peaks, Fe2+/Fe3+ mapping can be obtained using an electron probe microanalyzer.

Journal

References(23)*help

See more

Details 詳細情報について

Report a problem

Back to top