Sequential Determination of Calcium and Magnesium Cations in Haemodialysis Solutions by FIA.

  • LAPA Rui A. S.
    CEQUP/Departamento de Química Física, Faculdade de Farmácia, Universidade do Porto
  • LIMA José L. F. C.
    CEQUP/Departamento de Química Física, Faculdade de Farmácia, Universidade do Porto
  • VELA M. H.
    Departamento de Química Analítica, Facultad de Ciencias, Universidad de Valladolid
  • ARRADO E.
    Departamento de Química Analítica, Facultad de Ciencias, Universidad de Valladolid

書誌事項

タイトル別名
  • Sequential Determination of Calcium and

この論文をさがす

抄録

This work reports on the development, optimization and application of an FIA manifold to the sequential determination of calcium and magnesium cations in haemodialysis solutions. A tubular electrode sensitive to calcium and an atomic absorption spectrophotometer to measure magnesium were used as a detection processed. The system included a dialysis unit to enable sample dilution, since they presented high levels of the species to be determined. The system performance was optimized to obtain the maximum slope of calibration lines to Ca2+ and Mg2+ by means of the Taguchi Parameters Design (TPD). An L9(34) saturated orthogonal array led to the following optimized conditions: flow rate, 3.0ml/min; sample volume injected, 20μl; reactor length, 100cm; ionic strength, 3M. Under these conditions, the analytical procedure was robust and lead to reproducible results which were insensitive to variations in the glucose concentration. The developed system presents a high sampling rate (80 samples/h). The results obtained by the FIA system were compared with those given by the process of visual titration (10 samples/h) used as a reference method; it was found that both methods were in good agreement.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 13 (3), 409-414, 1997

    社団法人 日本分析化学会

参考文献 (29)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ