Characteristics of Sensor Response of a Difference Interferometric Slab Optical Waveguide Refractive Index Sensor with a Prism Coupling Method.
-
- TSUNODA Kin-ichi
- Department of Chemistry, Gunma University
-
- KIKUCHI Sae
- Department of Chemistry, Gunma University
-
- NOMURA Kiyohisa
- Department of Chemistry, Gunma University
-
- AIZAWA Katsuyuki
- Department of Chemistry, Gunma University
-
- OKAMOTO Kanae
- Department of Chemistry, Gunma University
-
- AKAIWA Hideo
- Department of Chemistry, Gunma University
Search this article
Abstract
We developed a difference interferometric slab optical waveguide (SOWG) refractive index (RI) sensor for flow analysis which used a prism coupling method. The highest sensitivity of 7.2×10-7 (ΔRI unit) was obtained with a glass SOWG fabricated with a radio frequency sputtering method for solution samples such as sucrose solutions and mixtures of ethanol and water; here sensitivity was defined as the ΔRI which gives the signal intensity of 10-4 in absorbance unit. The sensitivity is almost the same level as that of a surface plasmon resonance sensor.
Journal
-
- Analytical Sciences
-
Analytical Sciences 15 (3), 241-247, 1999
The Japan Society for Analytical Chemistry
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390001204253239808
-
- NII Article ID
- 130004440025
- 10002419033
-
- NII Book ID
- AA10500785
-
- COI
- 1:CAS:528:DyaK1MXhvVOgtLs%3D
-
- ISSN
- 13482246
- 09106340
-
- NDL BIB ID
- 4810234
-
- Text Lang
- en
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed