PET装置のQA/QC Quality assurance and quality control for positron emission tomographs

この論文にアクセスする

この論文をさがす

著者

    • 織田 圭一 ODA Keiichi
    • 東京都老人総合研究所ポジトロン医学研究部門 Positron Medical Center, Tokyo Metropolitan Institute of Gerontology

抄録

Quality assurance (QA) and quality control (QC) system is essential to positron emission tomographs (PET) like any other medical apparatus. Although part of QA/QC protocols are established by Japan Industries Association of Radiation Apparatus (JIRA), it is desirable to establish a comprehensive QA system that can cover a wide range of fields including physicists, chemists, medical doctors and technologists.<BR>The performance evaluation for PET QA/QC is standardized by National Electrical Manufacturers Association (NEMA), and an international standard of performance evaluation is under discussion by International Electrotechnical Commission (IEC). In Japan, a guideline for positron emission tomographs was published by Japan Radioisotope Association (JRIA) and standardized by JIRA. But there are some differences in methods of measurements and analyses among those publications.<BR>There are also QA/QC items which are not handled in the above mentioned guidelines, for example, check of detectors and cross calibration between PET and well counter.

収録刊行物

  • 放射線医学物理

    放射線医学物理 17(1), 37-43, 1997-03

    Japan Society of Medical Physics

参考文献:  16件中 1-16件 を表示

被引用文献:  1件中 1-1件 を表示

各種コード

  • NII論文ID(NAID)
    10003723911
  • NII書誌ID(NCID)
    AA11137895
  • 本文言語コード
    JPN
  • 資料種別
    ART
  • ISSN
    09188010
  • データ提供元
    CJP書誌  CJP引用  J-STAGE 