書誌事項
- タイトル別名
-
- Measurement of Absorbed Dose by Solid Calorimeter
- (1) Design and Basic Characteristics of a Solid Carbon Calorimeter
- 1)固体カーボン熱量計の試作と基礎特性
この論文をさがす
抄録
To determine an absorbed dose for proton and heavy ion beams, a quasi adiabatic solid calorimeter with carbon absorber was designed and constructed. The absorber is a graphite block with 2.5cm diameter and 4mm thick. It has contained three thermistors, one is used as a detector for the absorbed energy and the another two as the electrical temperature calibrators. We confirmed the system had a sufficient reliability to make the measurement of absorbed dose for dose rate larger than 0.5Gy/min.
収録刊行物
-
- Japanese Journal of Medical Physics
-
Japanese Journal of Medical Physics 18 (4), 327-332, 1998
一般社団法人 日本医学物理学会