Chemiluminescence Flow Biosensor System for Cholesterol with Immobilized Reagents.

  • HUANG Yuming
    Department of Chemistry, Nanjing University
  • ZHANG Chao
    Department of Chemistry, Shaanxi Normal University
  • ZHANG Zhujun
    Department of Chemistry, Shaanxi Normal University Department of Chemistry, Southwest Normal University

この論文をさがす

抄録

A chemiluminescence biosensor combined with flow injection analysis has been developed for determining cholesterol. Cholesterol oxidase was immobilized onto amine-modified silica gel via glutaraldehyde activation and packed in a column. The analytical reagents, including luminol and ferricyanide, were electrostatically coimmobilized on an anion-exchange column. Cholesterol was sensed by the CL reaction between hydrogen peroxide released from the enzymatic reaction and luminol and ferricyanide, which were released from a column with immobilized reagents by elution. The calibration graph was linear over the range 5×10-6 to 1×10-4 g ml-1 and the detection limit was 5×10-6 g ml-1. The pro-posed method has been successfully applied to the determination of cholesterol in human serum.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 15 (9), 867-870, 1999