人工血管の術後口径拡張に関する臨床的評価法 ノギスを用いた術中口径計測法の応用

書誌事項

タイトル別名
  • Postoperative dilatation of gelatin-sealed knitted Dacron prostheses: An intraoperative evaluation with slide calipers.
  • An intraoperative evaluation with slide calipers
  • ノギスを用いた術中口径計測法の応用

この論文をさがす

抄録

Knitted Dacron prostheses are known to dilate after surgery because of its configuration. Dilatation of gelatin-sealed knitted Dacron prosthesis (Gelseal®, Vascutek, UK) was evaluated in 9 patients who underwent surgical repair of abdominal aortic aneurysm. Prior to the implantation, the internal and external diameter of Gelseal were measured with slide calipers. Immediately after releasing the aortic clamp, the external diameter was also measured in the same fashion. Four weeks later, the external diameter was examined by CT scan. The size of Gelseal averaged 18.7±2.1mm. The internal and external diameters obtained from slide calipers measurements prior to the implantation were 19.2±2.2mm and 22.6±2.3mm. Immediately after the implantation, the external diameter revealed 22.7±2.4mm. Four weeks later, the external diameter averaged to 24.0±2.2mm. In conclusion, actual dilatation of the external diameter for one month following the implantation was 1.3mm (5.7%).

収録刊行物

  • 人工臓器

    人工臓器 24 (5), 1008-1010, 1995

    一般社団法人 日本人工臓器学会

被引用文献 (2)*注記

もっと見る

参考文献 (8)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘