書誌事項
- タイトル別名
-
- Development of motor-driven implantable left ventricular assist system: Basic design and its evaluation.
- ―基礎的検討―
この論文をさがす
抄録
We have developed a motor-driven implantable left ventricular assist system. This system consists of a pusher-plate type pump and a spline actuator driven by a DC brushless motor. A pump housing was made of epoxy-resin and a diaphragm was made of segmented- polyurethane (TM-5). The diameter of the model 1 was 85mm, 76mm in depth and 900 gram of weight. The actuator showed good durability (up to 300 days) in unloading status. In mock circulation, the model 1 showed 2.6l/min of output in 20mmHg of preload and 80mmHg of afterload. We changed the configuration of the pump and the motor design in model 2. New model showed 7.6l/min of output in mock circulation. This system is promising to the new implantable LVAS.