書誌事項
- タイトル別名
-
- The cleaning and Disinfection of the Hemodialysis Equipment Using Electrolysed Hyperacidity Water.
この論文をさがす
抄録
From November 1 994 to August 1995, we used electrolysed hyperacidity water (EHW) to clean and disinfect hemodialysis equipment and several pipe lines. In order to clarify the effectiveness of EHW cleaning and disinfection, we carried out a comparative study with conventional disinfectant agents, sodium hypochlorite, acetic acid and formalin. The effectivenss of sterilization using EHW and the decrease of endotoxin level were examined during the study. The effect of EHW disinfection appears immediately after using it and shows relatively high. In addition, EHW is more safe and easy to handle in comparison with other agents. Shorter rinse cycle and smaller amout usage in BHW are possible without other agents. Furthermore, the prevention of carbonate precipitation have become possible. Also, erosion caused by chloride gas has not been observed. Summary, EHW has many advantages to conventional disinfectant agents. An automated disinfection system using EHWwill be effective for dialysat purification.
収録刊行物
-
- 人工臓器
-
人工臓器 25 (2), 393-398, 1996
一般社団法人 日本人工臓器学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282679932151552
-
- NII論文ID
- 10007718045
-
- NII書誌ID
- AN00120167
-
- ISSN
- 18836097
- 03000818
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可