下顎運動測定装置CONDYLOCOMP LR3の測定精度 The Measurement Accuracy of Mandibular Movement Measuring Device "CONDYLOCOMP LR 3"

この論文にアクセスする

この論文をさがす

著者

    • 荒井 眞喜 ARAI Maki
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo
    • 北村 学 KITAMURA Manabu
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo
    • 林 幸男 HAYASHI Yukio
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo
    • 大山 茂稔 OHYAMA Shigetoshi
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo
    • 中田 浩史 NAKADA Hiroshi
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo
    • 小見山 道 KOMIYAMA Osamu
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo
    • 川良 美佐雄 KAWARA Misao
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo
    • 小林 喜平 KOBAYASHI Kihei
    • 日本大学松戸歯学部補綴学第I講座 Department of Complete Denture Prosthodontics, Nihon University of Dentistry at Matsudo

抄録

CONDYLOCOMP LR 3 is mandibular movement measuring device characterized by six degrees of freedom and a light reflective mechanism. The measurement accuracy of this device was evaluated.<BR>The results were as follows:<BR>1. The sagittal plane was expanded 4.8% in the sagittal and 5.6% in the vertical directions, by moving the reflector anteroinferiorly (the measuring point was 2.5 mm anterior and 2.5 mm inferior to the coordinate point). Further, the measuring point was 5.0 mm anterior and 5.0 mm inferior to the coordinate point, and expansion was 3.6% in the sagittal and 4.0% in the vertical directions.<BR>2. The sagittal plane was expanded 7.2% in the sagittal and 3.2% in the vertical directions, by moving the reflector internally 2.5 mm and anteroinferiorly (the measuring point was 2.5 mm anterior and 2.5 mm inferior to the coordinate point). Further, the measuring point was 5.0 mm anterior and 5.0 mm inferior to the coordinate point, and expansion was 7.4% in the sagittal and 4.3% in the vertical directions.<BR>3. The horizontal plane was expanded 7.2% in the sagittal and reduced 9.2% in the horizontal directions, by moving the reflector internally 2.5 mm and anteroinferiorly (the measuring point was 2.5 mm anterior and 2.5 mm internally to the coordinate point). Further, the measuring point was 5.0 mm anterior and 5.0 mm internally to the coordinate point, and expansion was 10.2% in the sagittal and reduced 3.2% in the horizontal directions.

収録刊行物

  • 日本補綴歯科學會雜誌 = The journal of the Japan Prosthodontic Society

    日本補綴歯科學會雜誌 = The journal of the Japan Prosthodontic Society 42(3), 431-437, 1998-06-10

    Japan Prosthodontic Society

参考文献:  16件中 1-16件 を表示

被引用文献:  4件中 1-4件 を表示

各種コード

  • NII論文ID(NAID)
    10008169753
  • NII書誌ID(NCID)
    AN00197911
  • 本文言語コード
    JPN
  • 資料種別
    ART
  • ISSN
    03895386
  • データ提供元
    CJP書誌  CJP引用  J-STAGE 
ページトップへ