Performance of a Prototype of Liquid Xenon Scintillation Detector System for Positron Emission Tomography

  • Nishikido Fumihiko
    Kikui-Branch, Advanced Research Institute for Science and Engineering, Waseda University
  • Doke Tadayoshi
    Kikui-Branch, Advanced Research Institute for Science and Engineering, Waseda University
  • Kikuchi Jun
    Kikui-Branch, Advanced Research Institute for Science and Engineering, Waseda University
  • Mori Toshinori
    Kikui-Branch, Advanced Research Institute for Science and Engineering, Waseda University
  • Takizawa Kyoko
    Kikui-Branch, Advanced Research Institute for Science and Engineering, Waseda University
  • Yamamoto Mikio
    Kikui-Branch, Advanced Research Institute for Science and Engineering, Waseda University

Search this article

Abstract

Liquid Xe as a scintillator has a very fast response. The scintillation light decays with 2.2 ns and 22 ns, and the recombination light for gamma-rays decays with 45 ns. Also, the scintillation yield is very high, comparable to that of NaI(Tl) crystals. These properties are considered to be advantageous for application to time of flight-positron emission tomography (TOF-PET). From this perspective, a prototype for a TOF-PET system which consists of two identical liquid xenon detectors has been constructed using liquid Xe as scintillator material and its energy resolution, position resolution and the time resolution were measured. The energy resolutions of 29.9% and 32.4% and position resolutions of 4.8 mm and 9.4 mm were obtained for the two identical detectors, respectively, and a time resolution of 650 ps between the two detectors was obtained for 511 keV gamma rays from positron annihilation. The future development of liquid Xe TOF-PET is discussed.

Journal

Citations (2)*help

See more

References(10)*help

See more

Details 詳細情報について

Report a problem

Back to top