脳磁図における三次元磁気計測 A Three-Dimensional Magnetic Field Measurement for Magnetoencephalogram

この論文にアクセスする

この論文をさがす

著者

抄録

Biomagnetic field measurement is widely used the magnetic field perpendicular to the scalp. There are problems of separating multiple sources overlapping in time when many distinct areas are active. We developed a three-dimensional (3-D) second-order gradiometer connected to three SQUIDs for vector measurement of the MEG (magnetoencephalogram). This SQUID magnetometer can detect magnetic field components perpendicular (Br) and tangential (Bθ, Bφ) to the scalp simultaneously. Magnetic field distribution of tangential component can provide information about new constraint conditions for inverse problem by visual inspection. This paper presents example of a 3-D vector measurement of MEG.

収録刊行物

  • 電気学会論文誌. E, センサ・マイクロマシン準部門誌 = The transactions of the Institute of Electrical Engineers of Japan. A publication of Sensors and Micromachines Society

    電気学会論文誌. E, センサ・マイクロマシン準部門誌 = The transactions of the Institute of Electrical Engineers of Japan. A publication of Sensors and Micromachines Society 124(6), 199-202, 2004-06-01

    一般社団法人 電気学会