Determination of Oxine Copper in Environmental Water by Activated Carbon Column Extraction / HPLC-UV

この論文にアクセスする

この論文をさがす

著者

抄録

A simple and low cost means of determining the concentration of oxine copper in tap and river water by high performance liquid chromatography (HPLC) coupled to an ultraviolet detector (UV) has been investigated. A good linearity of the calibration curve was obtained in the concentration range of 0.1 mgl<SUP>-1</SUP> to 2 mgl<SUP>-1</SUP> (r=0.999-0.988) . The detection limit based on <I>S/N</I>=3 was 0.075, μg ml<SUP>-1</SUP> for oxine copper. Acetonitrile was recommended to elute oxine copper, based on method development for the cartridge-type solid phase. The recovery for oxine copper from river water spiked at a concentration of 5 μgl<SUP>-1</SUP> was 88.8%, and the relative standard deviation was 3.3 % using the activated carbon column method. This analytical method was successfully applied to the determination of oxine copper in tap and river water.

収録刊行物

  • 環境化学 : journal of environmental chemistry

    環境化学 : journal of environmental chemistry 14(3), 643-648, 2004-09-22

    一般社団法人 日本環境化学会