Constructing a New Optical Sensor for Monitoring Ammonia in Water Samples Using Bis(acetylacetoneethylendiamine)-tributylphosphin Cobalt(III) Tetraphenylborate Complex-Coated Triacetylcellulose

この論文にアクセスする

この論文をさがす

著者

抄録

A new ammonia optical sensor was designed using bis(acetylacetoneethylendiamine)tributylphosphin cobalt(III) tetraphenylborate complex, coated on transparent triacetylcellulose film as membrane. The change in the absorbance of the optode at the maximum wavelength of 408 nm was related to ammonia concentration in aqueous samples. A buffer solution with a pH of 9 (sodium borate-HCl) was used. The optode was fully regenerated in pH 2. The linear dynamic range for determination of ammonia was 3.3 × 10<sup>-4</sup> to 6.9 × 10<sup>-3</sup> mol l<sup>-1</sup> with a detection limit of 5.0 × 10<sup>-5</sup> mol l<sup>-1</sup> and a response time range of 4 - 6 min. This membrane was successfully applied for determination of ammonia in drinking water.

収録刊行物

  • Analytical sciences : the international journal of the Japan Society for Analytical Chemistry

    Analytical sciences : the international journal of the Japan Society for Analytical Chemistry 20(10), 1433-1436, 2004-10-10

    社団法人 日本分析化学会

参考文献:  40件中 1-40件 を表示

各種コード

  • NII論文ID(NAID)
    10013657489
  • NII書誌ID(NCID)
    AA10500785
  • 本文言語コード
    ENG
  • 資料種別
    ART
  • ISSN
    09106340
  • NDL 記事登録ID
    7111419
  • NDL 雑誌分類
    ZP4(科学技術--化学・化学工業--分析化学)
  • NDL 請求記号
    Z54-F482
  • データ提供元
    CJP書誌  NDL  J-STAGE 