キセノンCT検査の頭部固定法に関する検討―陰圧式固定具の有用性―

書誌事項

タイトル別名
  • Head holder using negative pressure bag packed with plastic beads in xenon CT CBF study
  • 陰圧式固定具の有用性

この論文をさがす

抄録

Employing analysis of CBF confidence maps, we investigated the usefulness of a head holder using a negative pressure bag packed with plastic beads in a xenon CT CBF study. A total of 272 consecutive patients for the CBF study were enrolled and classified into 3 groups : 88 patients with a negative pressure bag (M group), 87 patients with an air pillow (A group), and 97 patients with a sponge pillow (S group). The degree of effect of head movements on the CBF measurement in each patient was expressed as a confidence value (mean of the confidence values at one CT slice). The mean of confidence value in the M group (0.461) was statistically lower than that in the A group (0.866) and that in the S group (1.043). These findings showed that the head holder described here was useful for obtaining CBF maps of high quality in a xenon CT CBF study.

収録刊行物

  • 脳卒中

    脳卒中 25 (2), 278-281, 2003