Head holder using negative pressure bag packed with plastic beads in xenon CT CBF study
-
- Araki Yuzo
- Department of Neurosurgery, Inuyama Central Hospital
-
- Sakai Noboru
- Department of Neurosurgery, Gifu University
Bibliographic Information
- Other Title
-
- キセノンCT検査の頭部固定法に関する検討―陰圧式固定具の有用性―
- 陰圧式固定具の有用性
Search this article
Abstract
Employing analysis of CBF confidence maps, we investigated the usefulness of a head holder using a negative pressure bag packed with plastic beads in a xenon CT CBF study. A total of 272 consecutive patients for the CBF study were enrolled and classified into 3 groups : 88 patients with a negative pressure bag (M group), 87 patients with an air pillow (A group), and 97 patients with a sponge pillow (S group). The degree of effect of head movements on the CBF measurement in each patient was expressed as a confidence value (mean of the confidence values at one CT slice). The mean of confidence value in the M group (0.461) was statistically lower than that in the A group (0.866) and that in the S group (1.043). These findings showed that the head holder described here was useful for obtaining CBF maps of high quality in a xenon CT CBF study.
Journal
-
- Japanese Journal of Stroke
-
Japanese Journal of Stroke 25 (2), 278-281, 2003
The Japan Stroke Society