Application of an in-situ Thermo-polymerized Porous Polymer: Creation of an On-column Frit for a Packed Capillary HPLC Column
-
- MA Jiping
- Institute of Environment & Municipal Engineering, Qingdao Technological University Key Lab of Bioorganic Phosphorus Chemistry & Chemical Biology, Ministry of Education, Department of Chemistry, Tsinghua University
-
- DING Mingyu
- Key Lab of Bioorganic Phosphorus Chemistry & Chemical Biology, Ministry of Education, Department of Chemistry, Tsinghua University
-
- XU Yan
- School of the Earth Sciences and Resources, China University of Geosciences
-
- CHEN Lingxin
- Key Lab of Bioorganic Phosphorus Chemistry & Chemical Biology, Ministry of Education, Department of Chemistry, Tsinghua University
この論文をさがす
抄録
A 3-mm length of a porous monolithic polymer was prepared in a 0.32-mm inner-diameter fused-silica capillary by an in-situ thermo-polymerization method and used as an on-column frit for a packed capillary HPLC column. The on-column frit can resist high pressure up to 400 bar. A 5-µm packing material was packed in the capillary with the on-column frit by a slurry method. At pressure driving mode, separation of samples was performed using the capillary HPLC column. The in-situ frit preparation method has the advantages of easy preparation, easy control of the location of the frit and a mild preparing reaction condition.
収録刊行物
-
- Analytical Sciences
-
Analytical Sciences 23 (3), 371-374, 2007
社団法人 日本分析化学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390001204256442368
-
- NII論文ID
- 130004441435
- 10018824280
-
- NII書誌ID
- AA10500785
-
- COI
- 1:CAS:528:DC%2BD2sXjtlekuro%3D
-
- ISSN
- 13482246
- 09106340
-
- NDL書誌ID
- 8672123
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可