Special issue on "silicon chemical sensors": Automation of pH measurement using a flow-through type differential pH sensor system based on pH-FET
-
- Mohri Satoshi
- Department of Cardiovascular Physiology, Okayama University Graduate School of Medicine, Dentistry, and Pharmaceutical Sciences
-
- Nakamura Michihiro
- Department of Cardiovascular Physiology, Okayama University Graduate School of Medicine, Dentistry, and Pharmaceutical Sciences
-
- Naruse Keiji
- Department of Cardiovascular Physiology, Okayama University Graduate School of Medicine, Dentistry, and Pharmaceutical Sciences
書誌事項
- タイトル別名
-
- Automation of pH Measurement Using a Flow-Through Type Differential pH Sensor System Based on pH-FET
この論文をさがす
抄録
Automatic pH measurement system was developed by using a flow-through type differential pH-FET sensor. The system was operated by repeating 60 sec cycle consisting of 36 sec of washing period, 4 sec of moving period, and 20 sec of measuring period. In washing period, the differential pH-FET sensor was washed with baseline solution. In measuring period, about 50 μl of sample solution was withdrawn to the space around a measurement pH-FET, and the change of differential source potential was measured. Absolute values of error for nine sample solutions with weak buffering ability were less than 0.146 pH, and were 0.071 pH in average. From this experiment, it was suggested that high-speed and automated pH measurement will be possible by using pH-FET sensors even for sample solutions of weak buffer ability.
収録刊行物
-
- 電気学会論文誌E(センサ・マイクロマシン部門誌)
-
電気学会論文誌E(センサ・マイクロマシン部門誌) 127 (8), 367-370, 2007