Optimization of Conditions for High-Performance Size-Exclusion Chromatography of Different Soil Humic Acids

Bibliographic Information

Other Title
  • Development and validation of method of high-performance size-exclusion chromatography (HPSEC) for various soil humic acids

Search this article

Abstract

A method of high-performance size-exclusion chromatography (HPSEC) for a wide variety of soil humic acids (HAs) was developed. Two types of soil HAs (Cambisol and Andosol HAs), which have substantially different chemical properties, showed different effects of salt and organic solvent concentrations in the eluent on chromatograms. A Shodex OHpak SB-805 HQ column with 10 mM sodium phosphate buffer (pH 7.0) containing 25% of acetonitrile (v/v) was found to be applicable for different HAs, and showed high reproducibility and recovery (87.0 - 94.5%). The Cambisol HA was fractionated into five fractions using an ultrafiltration with different molecular-weight cut-offs. The order of the molecular weights of the five fractions calculated from the HPSEC analysis corresponded to that defined by ultrafiltration. This supported the reliability of the method.

Journal

  • Analytical Sciences

    Analytical Sciences 24 (5), 607-613, 2008

    The Japan Society for Analytical Chemistry

Citations (7)*help

See more

References(67)*help

See more

Details 詳細情報について

Report a problem