Factors That Influence Measurements and Accurate Evaluation of Stent Apposition by Optical Coherence Tomography Assessment Using a Phantom Model

  • Sawada Takahiro
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Shite Junya
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Negi Noriyuki
    Department of Radiological Technology, Kobe University Hospital
  • Shinke Toshiro
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Tanino Yusuke
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Ogasawara Daisuke
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Kawamori Hiroyuki
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Kato Hiroki
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Miyoshi Naoki
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Yoshino Naoki
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Kozuki Amane
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine
  • Koto Masanobu
    Department of Radiological Technology, Kobe University Hospital
  • Hirata Ken-ichi
    Division of Cardiovascular Medicine, Department of Internal Medicine, Kobe University Graduate School of Medicine

Bibliographic Information

Other Title
  • Assessment Using a Phantom Model

Search this article

Abstract

Background: Factors influencing measurements and accurate evaluation of stent apposition by optical coherence tomography (OCT) are not established. Methods and Results: Phantom models of known luminal sizes and lengths were evaluated by OCT under various conditions and measurements were compared with actual values. Stents implanted into phantom models were examined by OCT to validate the measurement point on the stent strut surface for accurate evaluation of stent apposition. Strut thickness was measured at 3 points (midpoint, inner and outer surfaces of the stent shadow). The precision of OCT measurements of lumen diameter and area was satisfactory when the image wire was positioned in the center, but the error and deviation were unsatisfactory when the image wire was positioned eccentrically using a low frame acquisition rate. Longitudinal OCT measurements were close to actual values under all conditions examined. Measurements from the midpoint of the stent shadow to the adjacent vessel wall surface coincided with actual stent thickness. Conclusions: Significant measurement error can occur if the image wire is positioned eccentrically with a lower frame acquisition rate than specified by the manufacturer. To accurately evaluate stent apposition, the stent surface should be measured from the center of the stent reflection. (Circ J 2009; 73: 1841-1847)<br>

Journal

  • Circulation Journal

    Circulation Journal 73 (10), 1841-1847, 2009

    The Japanese Circulation Society