Development of the Conduction Calorimeter with Heat Sink Controlled Directly Using Peltier Device

Bibliographic Information

Other Title
  • ペルティエ素子によるヒートシンク直接制御方式の伝導熱量計の開発
  • ペルティエ ソシ ニ ヨル ヒートシンク チョクセツ セイギョ ホウシキ ノ デンドウ ネツリョウケイ ノ カイハツ

Search this article

Abstract

A conduction calorimeter with a thermally stable heat sink and six measurement cells have been designed and constructed. In order to make a compact calorimeter which is easy to operate in application to quality control of industrial materials such as cement, temperature of the aluminum heat sink was controlled directly using Peltier effect of Thermomodule. Temperature fluctuation of the heat sink was within ±0.1 ℃ during 8 days of continuous measurement accompanying large exothermic heat of reaction. Les than 40 μV/24h(1.38×10 -3℃)and 20 μVp-p(±0.69×10 -3℃)were achieved for base line drift and thermal noise, respectively. Both accuracy and reproducibility were estimated to be better than 0.5 %. Interference between the neighboring cells was negligible in simultaneous measurement of six samples. It was confirmed that the developed calorimeter was successfully applicable to measurement of industrial materials with extremely low reaction rate.

Journal

  • Netsu Sokutei

    Netsu Sokutei 38 (3), 83-85, 2011-06-30

    The Japan Society of Calorimetry and Thermal Analysis

References(4)*help

See more

Related Projects

See more

Details 詳細情報について

Report a problem

Back to top