放射線治療計画装置の MLC Leaf Offset Table の最適化に関する検討 Optimization of Multi-leaf Collimator Leaf Offset Table for a Treatment Planning System

この論文にアクセスする

この論文をさがす

著者

抄録

For accurate dose delivery using a rounded leaf-end multi-leaf collimator (MLC), a consistent understanding of MLC leaf offset is required between a linear accelerator (linac) controller and a treatment planning system (TPS). We have found that the leaf offset tables were inconsistent between an Elekta linac controller and a Pinnacle TPS. An optimum leaf offset table for Pinnacle was determined by operating the TPS and the linac controller. It was suggested that the MLC leaf transmission value and the MLC leaf offset table stored in Pinnacle may need to be revised to improve dose verification results for volumetric modulated arc therapy.

収録刊行物

  • 日本放射線技術學會雜誌

    日本放射線技術學會雜誌 68(2), 173-176, 2012-02-20

    公益社団法人 日本放射線技術学会

参考文献:  5件中 1-5件 を表示