A Small-Area and Low-Power SoC for Less-Invasive Pressure Sensing Capsules in Ambulatory Urodynamic Monitoring

この論文にアクセスする

この論文をさがす

著者

    • IWATO Hirofumi
    • the Graduate School of Information Science and Technology, Osaka University
    • SAKANUSHI Keishi
    • the Graduate School of Information Science and Technology, Osaka University
    • IMAI Masaharu
    • the Graduate School of Information Science and Technology, Osaka University

抄録

To measure the detrusor pressure for diagnosing lower urinary tract symptoms, we designed a small-area and low-power System on a Chip (SoC). The SoC should be small and low power because it is encapsulated in tiny air-tight capsules which are simultaneously inserted in the urinary bladder and rectum for several days. Since the SoC is also required to be programmable, we designed an Application Specific Instruction set Processor (ASIP) for pressure measurement and wireless communication, and implemented almost required functions on the ASIP. The SoC was fabricated using a 0.18µm CMOS mixed-signal process and the chip size is 2.5×2.5mm<sup>2</sup>. Evaluation results show that the power consumption of the SoC is 93.5µW, and that it can operate the capsule for seven days with a tiny battery.

収録刊行物

  • IEICE transactions on electronics

    IEICE transactions on electronics 95(4), 487-494, 2012-04-01

    一般社団法人 電子情報通信学会