46 植物ホルモンの分析のためのLC-MS-SIMの高感度の設定条件

DOI

書誌事項

タイトル別名
  • Established Conditions of Higher Sensitivity of LC-MS-SIM for phytohormone analysis

抄録

For a higher sensitivity, the parameters of liquid chromatography-mass spectrometry-selected ion monitoring (LC-MS-SIM) usirtg atmospheric pressure chemical ionization (APCI) were examined. The area and height of a peak as sensitivity, the full width half maximum (FWHM) and the actual scan-time which was required for the scanning were measured. The suitable parameters for higher sensitivity on measurement limit level were as follows: 1) The scan-tlme might correspond to the value of FWHM. 2) The countling-number was over about 7,000. 3) The scan-range might correspond to a decimal fraction of the molecular weight of the sample substance. The results showed that the value obtained from repeated sampling-point-number was a total amount, that the value obtained from repeated counting-number was a mean.

収録刊行物

詳細情報 詳細情報について

  • CRID
    1390282679400973056
  • NII論文ID
    110001853821
  • DOI
    10.18978/jscrpanb.36.0_103
  • ISSN
    21896313
    09191887
  • 本文言語コード
    ja
  • データソース種別
    • JaLC
    • CiNii Articles
  • 抄録ライセンスフラグ
    使用不可

問題の指摘

ページトップへ