電気化学検出高速液体クロマトグラフィーによる血漿中微量ビタミンK類測定法の検討

書誌事項

タイトル別名
  • Determination of Vitamin K in Plasma by High Performance Liquid Chromatography with Electrochemical Detection
  • デンキ カガク ケンシュツ コウソク エキタイ クロマトグラフィー ニ ヨル

この論文をさがす

抄録

Although the methods of high-performance liquid chromatography with electrochemical detection for the determination of vitamin K (K_s) in human plasma had already reported, they have some problems including the effect of dissolved oxygen in mobile phase on background current, losing sensitivity of electrode due to adsorption of reducted compounds on to the surface of the electrode, and single determination for phylloquinone (K_1). For the resolution of these problems, we tried to use a platinum black column for the reduction of K_s instead of electrochemical reduction, and an amperometric electrochemical detector in the oxidation mode. Therefore, we developed a sensitive method for the simultaneous determination of K_1 and menaquinons (MK-4, MK-7) in human plasma. The detection limits in the present method were 0.05 ng/ml of K_1, MK-4 and 0.1 ng/ml of MK-7.

収録刊行物

  • ビタミン

    ビタミン 68 (12), 711-719, 1994

    公益社団法人 日本ビタミン学会

被引用文献 (3)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘