Round robin test on dielectric resonator method for measuring complex permittivity at microwave frequency

Search this Article

Author(s)

Abstract

The dielectric resonator method is now widely accepted as a precise measurement method for determining the dielectric properties at microwave frequencies. This paper describes the measurement results of ε_r, tan δ and TC_f determined by a round robin test of this method. The resultant measurement errors were Δε_r/ε_r=0.10%, Δtanδ<0.5×10^<-5> and ΔTC_f<0.5ppm/K, where Δ denotes a standard deviation. The causes of measurement errors and the conditions to improve the measurement accuracy are discussed.

Journal

  • IEICE Trans. Electron., C

    IEICE Trans. Electron., C E77-C(6), 882-887, 1994

    The Institute of Electronics, Information and Communication Engineers

Cited by:  21

Codes

  • NII Article ID (NAID)
    110003220420
  • NII NACSIS-CAT ID (NCID)
    AA10826283
  • Text Lang
    ENG
  • Article Type
    Journal Article
  • ISSN
    09168524
  • Data Source
    CJPref  NII-ELS 