乳房撮影装置の半価層測定に関する検討 Measurement of Half-value Layer in Mammography

この論文にアクセスする

この論文をさがす

著者

    • 八木 浩史 YAGI HIROFUMI
    • 徳島大学医学部保健学科放射線技術科学専攻 Department of Radiologic Science and Engineering, School of Health Sciences, The University of Tokushima
    • 北村 怜子 KITAMURA REIKO
    • 岡山大学医学部附属病院核医学診療室 Clinical Department of Nuclear Medicine, Okayama University Hospital
    • 土居 奈津子 DOI NATSUKO
    • 徳島大学医療技術短期大学部診療放射線技術学科:(現)高知医科大学医学部付属病院放射線部 Department of Radiological Technology, School of Medical Sciences, The University of Tokushima:(Present address)Department of Radiology, Kochi Medical School Hospital
    • 山根 英里子 YAMANE ERIKO
    • 徳島大学医療技術短期大学部診療放射線技術学科:(現)医療福祉センター倉吉病院放射線科 Department of Radiological Technology, School of Medical Sciences, The University of Tokushima(Present address)Department of Radiology, Kurayoshi Hospital

抄録

The half-value layer (HVL) of an X-ray beam for film-screen mammography is considered an important parameter for image quality and patient dose. Thus, HVL must be measured in accordance with The Manual of Accuracy for Mammography printed by the Japanese Society of Radiological Technology. The manual prescribes exactly the geometry of measurement, chamber position of measurement in the field, selection of chamber, and so on. However, the measurement of HVL is difficult in the actual clinical setting. This study examined the results of failure to perform the measurement of HVL in accordance with the manual for measuring HVL in the clinical setting. The investigation indicated that serious problems do not arise when measuring HVL for routine quality control even if the chamber in the field is not always set according to the manual and if a chamber for radiotherapy or diagnosis is used that is not recommended for soft X-ray by the manual.

収録刊行物

  • 日本放射線技術学会雑誌

    日本放射線技術学会雑誌 59(6), 729-736, 2003

    公益社団法人 日本放射線技術学会

参考文献:  17件中 1-17件 を表示

被引用文献:  5件中 1-5件 を表示

各種コード

  • NII論文ID(NAID)
    110003461289
  • NII書誌ID(NCID)
    AN00197784
  • 本文言語コード
    JPN
  • 資料種別
    ART
  • ISSN
    0369-4305
  • データ提供元
    CJP書誌  CJP引用  NII-ELS  J-STAGE 