Driven Equilibrium Pulse を利用した Turbo Spin Echo 法(DRIVE)の基礎的検討  [in Japanese] Fundamental Study of Turbo Spin Echo Sequence with Driven Equilibrium Pulse  [in Japanese]

Access this Article

Author(s)

Abstract

When MR images are obtained with the turbo spin echo (TSE) sequence, DRIVE can be used as a sequence in which the driven equilibrium pulse (DE pulse), a reset pulse, is applied at the TSE echo train to accelerate relaxation time and return to the equilibrium of Mz magnetization. In this study, we examined the extent to which TR could be shortened in DRIVE and how the other parameters of the turbo spin echo sequence influence it. Results: 1) DRIVE is effective when the T2 value is long. 2) It is necessary to set TR at 1000 ms or more to obtain image contrast with free water and fat in T2-weighted images for which a conventional turbo spin echo sequence using DRIVE is employed in clinical examination. 3) It is not necessary to consider the influence of the TSE factor when using DRIVE.

Journal

  • Japanese Journal of Radiological Technology

    Japanese Journal of Radiological Technology 62(7), 1013-1017, 2006-07-20

    Japanese Society of Radiological Technology (JSRT)

References:  9

Codes

  • NII Article ID (NAID)
    110004776027
  • NII NACSIS-CAT ID (NCID)
    AN00197784
  • Text Lang
    JPN
  • Article Type
    NOT
  • ISSN
    03694305
  • NDL Article ID
    8044691
  • NDL Source Classification
    ZS45(科学技術--医学--放射線医学)
  • NDL Call No.
    Z19-205
  • Data Source
    CJP  NDL  NII-ELS  J-STAGE 