ビジュアルサーボを用いたタッチ式内視鏡手術器具インタフェース

書誌事項

タイトル別名
  • Endoscopic Instrument Interface with Touch Panel Using Visual Servoing
  • ビジュアル サーボ オ モチイタ タッチシキ ナイシキョウ シュジュツ キグ インタフェース

この論文をさがす

抄録

To solve endoscopic surgical problems such as lack of depth information and hand-eye coordination, this paper proposes an intuitive human-machine interface for 3-D positioning of surgical instruments that uses the monitor as an input device of the desired point. This interface must achieve 3-D positioning from only 2-D image information. To cope with these problems, we propose a new 3-D positioning method using a laser pointer. This method achieves 3-D positioning of a surgical instrument by determining and controlling the direction of insertion of the instrument based on visual servoing, followed by estimating the distance between the instrument and organs using projective invariants. To evaluate validity of this interface, we conducted an experiment in which the interface system makes trajectory tracking with keeping the instrument-organ distance constant. As a result, it was confirmed that a laser beam was irradiated along the desired trajectory in few gaps while keeping the distance less variant.

収録刊行物

  • 生体医工学

    生体医工学 46 (6), 595-605, 2008

    公益社団法人 日本生体医工学会

被引用文献 (1)*注記

もっと見る

参考文献 (19)*注記

もっと見る

関連プロジェクト

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ