When blood comes in contact with artificial systems it tends to clot more. Hence, blood clotting is the major issue in extracorporeal circulation. Due to this, the patients take regular anticoagulation medicine. The frequent use of such a medicine results to other consequences. Due to this, the optimum use of medicine is desirable. To address this issue, we have made an assessment of instrumentation systems based on electrical impedance measurement with the aim of developing a sensor to monitor the blood and detect the thrombogenic activities at the early stage. The experimental results obtained by monitoring the bovine blood are presented. It was understood that electrical resistivity can be a suitable parameter to detect embolus. And, change in capacitance was more specific to understand the thrombogenic process.