S022022 再生血管の動的培養において負荷される拍動力学刺激パターンの評価([S022-02]循環器系医療機器(2))

書誌事項

タイトル別名
  • S022022 Evaluation of Mechanical Stimuli Patterns on the Pulsatile Culture of Tissue Engineered Vascular Grafts

抄録

In these days,the number of the case of vascular disease is increasing.To overcome these cases,tissue engineered blood vessels(TEBV) are regarded as a superior method to artificial or transplanted graft,which are short of antithrombogenicity and biocompatibility.In tissue engineering,developing tissues by cell culture,dynamic culture accompanied medium circulation and mechanical stimuli loading are beneficial to tissue maturation.Thus,various bioreactors are developed,and proved that loading pulsatile mechanical stimuli as pressure,shear stress and strain will encourage the tissue maturation in mechanical strength and differentiation of multi potential cells.In this study,we developed the new bioreactor to control the pulsatile mechanical stimuli to investigate the effect on TEBV growth.The system was operated and pulsatile pressure and shear stress were measured.The stimuli patterns were manipulated arbitrarily.Our bioreactor provided the new experimental system that can investigate the effect by various mechanical stimuli patterns on TEBV growth.

収録刊行物

  • 年次大会

    年次大会 2013 (0), _S022022-1-_S022022-4, 2013

    一般社団法人 日本機械学会

詳細情報 詳細情報について