書誌事項
- タイトル別名
-
- F-2-3 Bending moment measurement using grasping forceps with sensors
抄録
In this work, we measured the bending moment using a grasping forceps with sensors. Grasping operations during the endoscopic surgery apply forces and moments to the grasped object. The forces and moments cause damage to the grasped object. Therefore, measurement of the forces and moments is necessary to prevent the damage by the over application. We have measured the forces by the grasping forceps with two MEMS (Micro Electro Mechanical Systems) triaxial force sensors and a potentiometer. In this paper, we measured the bending moment induced by rotating a rigid rod between the tips of the developed grasping forceps. Measured bending moment was linear to the rotation angle and 1.7 mN・m for 1° rotation.
収録刊行物
-
- IIP情報・知能・精密機器部門講演会講演論文集
-
IIP情報・知能・精密機器部門講演会講演論文集 2014 (0), _F-2-3-1_-_F-2-3-2_, 2014
一般社団法人 日本機械学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282680877702656
-
- NII論文ID
- 110009939180
-
- ISSN
- 24243140
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可