書誌事項
- タイトル別名
-
- 509 Observation of in vitro thrombosis behavior under shear flow field
抄録
In vitro antithrombogenic testing is a useful pre-evaluation for ex vivo test of a continuous flow pump as an artificial heart. In our previous study, we have quantitatively obtained the results that the increasing shear rate prevented the thrombus formation using a double-cylinder type rheometer. However, it is not yet cleared that the thrombosis procedures occurred at between the inner and the outer cylinders. This paper describes that the optical and direct observation of the in vitro thrombosis under shear flow fields. This results showed that the thrombus formation started from the surface of rotating outer cylinder made of glass. Thus, it is considered that in vitro thrombosis closely related to the not only blood coagulability and/or shear rate, but also chemical condition of blood-contacting surface.
収録刊行物
-
- 茨城講演会講演論文集
-
茨城講演会講演論文集 2014.22 (0), 15-16, 2014
一般社団法人 日本機械学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282680830394112
-
- NII論文ID
- 110009966308
-
- ISSN
- 24242683
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可