Dynamic viscoelastic properties and the age changes of long-term soft denture liners

機関リポジトリ 被引用文献8件 オープンアクセス

この論文をさがす

抄録

The dynamic viscoelastic properties of long-term soft denture liners were measured over a wide range of frequencies using a dynamic viscoelastometer based on a non-resonance-forced vibration principle. Changes in properties over a 3 yr period have also been monitored. One acrylic material, one fluoroelastomer, one heat cured silicone and one self-curing addition silicone were used. Complex dynamic tensile modulus (E*), tensile storage modulus (E′), tensile loss modulus (E″) and loss tangent (tan δ) were determined over the frequency range from 0.01 to 100 Hz on administration of a 0.27(null)train at 37°C. The dynamic viscoelasticity of the acrylic and fluoroelastomer products was more sensitive to changes in frequency than that of silicone products. The acrylic material and fluoroelastomer exhibited viscoelastic behaviour whilst silicones exhibited elastic behaviour. The silicone products remained unchanged after soaking for 3 yr whilst the acrylic and fluoroelastomer products underwent significant change.

収録刊行物

  • Biomaterials

    Biomaterials 21 (14), 1421-1427, 2000-07

    Elsevier

被引用文献 (8)*注記

もっと見る