スぺクトラスパン・プラズマ発光分光光度計による温泉中のホウ素の定量  [in Japanese] Determination of boron in thermal waters by emission spectrophotometry using "Spectraspan"  [in Japanese]

Access this Article

Search this Article

Abstract

A plasma emission spectrophotometer "Spectraspan" (low power d. c. plasma arc, operating on argon) was evaluated for boron determination in thermal waters. The influence of acids (hydrochloric, nitric and sulfuric acids) and several metallic ions (Na(+). K(+), Mg(2+). Ca(2+)) on emission intensity was studied. There was a linear relationship between emission intensity and boron content from 0 to 500 mg/l. Boron in thermal waters was easily determined by standard addition method. Precision. coefficient of variance and recoveries of known amount of boron added to the sample for 11 repricate analyses were 0.12μg/ml, 2.02%, 95.0-101.7%, respectively. Boron content of 27 thermal waters in the Sanin district was determined. and the highest B content in the sample waters were 8.8 mg/l (Tottori spa) in Tottori Prefecture and 14.6-25.0 mg/l (the thermal springs at the foot of Mt. Sanbe. a non-active quaternary volcano) in Shimane Prefecture. The relationships between B content and water temperature, B content and pH value were not recognized. Significantly positive correlationship was observed between Band Li contents.

Journal

  • 岡山大学温泉研究所報告

    岡山大学温泉研究所報告 (48), 35-41, 1979-03-25

    岡山大学温泉研究所

Codes

  • NII Article ID (NAID)
    120002311017
  • NII NACSIS-CAT ID (NCID)
    AN00032853
  • Text Lang
    JPN
  • Article Type
    departmental bulletin paper
  • ISSN
    0369-7142
  • Data Source