A thermocouple-based remote temperature controller of an electrically-floated sample for plasma CVD of nanocarbons with bias voltage

抄録

We report an accurate and easy-to operate instrument for the temperature control of an electrically floated sample using a thermocouple in direct contact with it. The signal was transmitted via 2.45 GHz WiFi. We measured and analyzed the discrepancy between the thermocouple and a radiation thermometer for the plasma CVD of carbon nanomaterials under sample bias. A successful protection method from abnormal discharge in the plasma is also provided. (C) 2017 Elsevier Ltd. All rights reserved.

収録刊行物

  • Measurement

    Measurement 102 244-248, 2017-05

    Elsevier

被引用文献 (2)*注記

もっと見る

参考文献 (24)*注記

もっと見る

関連プロジェクト

もっと見る

詳細情報 詳細情報について

問題の指摘