Impact of the Characteristic Impedance of Coaxial Lines on the Sensitivity of a 750-MHz Electronically Tunable EPR Resonator
Search this article
Abstract
A 750-MHz electronically tunable resonator was investigated in terms of the sensitivity of electron paramagnetic resonance (EPR) signal detection. The conversion efficiency of the radio-frequency magnetic field was calculated for resonators with 50- and 100-Omega coaxial coupling lines using three-dimensional (3D) microwave field and microwave circuit simulators. Based on the simulation results, two tunable resonators were physically constructed and compared in terms of EPR signal sensitivity using a nitroxyl radical solution. While the resonator with 100-Omega coaxial lines provided 14% greater signal intensity, its signal-to-noise ratio was lower than that of the resonator with 50-Omega lines. To demonstrate the capability of the constructed tunable resonator for EPR imaging experiments, a solution of nitroxyl radical and the leg of a tumor-bearing mouse were visualized.
Journal
-
- Applied magnetic resonance
-
Applied magnetic resonance 49 (8), 853-867, 2018-08
Springer
- Tweet
Details 詳細情報について
-
- CRID
- 1050001339062325504
-
- NII Article ID
- 120006709282
-
- NII Book ID
- AA10933363
-
- HANDLE
- 2115/75089
-
- ISSN
- 09379347
-
- Text Lang
- en
-
- Article Type
- journal article
-
- Data Source
-
- IRDB
- CiNii Articles