The best DC voltage in plant hormone analysis using LC / MS

Bibliographic Information

Other Title
  • LC/MSを用いた植物ホルモン分析における分析部直流(DC)電圧の報告
  • LC/MS オ モチイタ ショクブツ ホルモン ブンセキ ニ オケル ブンセキブ チョクリュウ(DC)デンアツ ノ ホウコク

Search this article

Abstract

Mass spectrums were analyzed as DC voltage to determine the optimum condition in quantitative analysis of phytohormones ; IAA (indore 3-acetic acid), ABA (abscisic acid) and GA1(gibberellic acid1) and GA4(gibberellic acid4). Optimal DC voltage differed in each standard substance and analysis mode. In IAA,optimal DC voltage was 65V in Negative Mode, in ABA, 20V in Positive Mode, and in GA1, 20V in Positive Mode. As GA4 was very a small amount of samples, optimal DC voltage was 65V or 80V in Negative Mode.