Development of catheter-type tactile sensor composed of polyvinylidene fluoride (PVDF) film

抄録

To achieve quantitative palpation in vivo, we developed a catheter-type tactile sensor composed of a polyvinylidene fluoride film for minimally invasive surgery. We evaluated the fundamental performance of the prototype sensor by a weight-drop test. We also measured the output of the prototype sensor as it was inserted into a blood vessel model with shapes mimicking lesions. The ø2-mm sensor passed easily into the blood vessel model with lesion-like shapes. Sensor outputs corresponded to the shape of the inner wall of the blood vessel model, making it possible to determine the position of a protrusion and the convexity interval of a rough surface by filtering and frequency analysis of the output.